Please report any queries concerning the funding data grouped in the sections named "Externally Awarded" or "Internally Disbursed" (shown on the profile page) to your Research Finance Administrator. Your can find your Research Finance Administrator at https://www.ucl.ac.uk/finance/research/rs-contacts.php by entering your department
Please report any queries concerning the student data shown on the profile page to:

Email: portico-services@ucl.ac.uk

Help Desk: http://www.ucl.ac.uk/ras/portico/helpdesk
Publication Detail
Augmentation of reality using an operating microscope for otolaryngology and neurosurgical guidance
  • Publication Type:
    Journal article
  • Publication Sub Type:
    Journal Article
  • Authors:
    Edwards PJ, Hawkes DJ, Hill DLG, Jewell D, Spink R, Strong A, Gleeson M
  • Publication date:
    01/01/1995
  • Pagination:
    172, 178
  • Journal:
    Computer Aided Surgery
  • Volume:
    1
  • Issue:
    3
  • Status:
    Published
  • Print ISSN:
    1092-9088
Abstract
The operating microscope is an integral part of many neurosurgery and otolaryngology procedures; the surgeon often uses the microscopic view for a large portion of the operation. Information from preoperative radiological images is often viewed only on X-ray films. The surgeon then has die difficult task of relating this information to the appearance of the surgical view. Image guidance techniques attempt to relate these two sets of information by registering the patient in the operating room to preoperative images using locating devices. Conventionally, image data are presented on a computer monitor, which requires the surgeon to look away from the operative scene. We describe a guidance system, for procedures in which the operating microscope is used, which superimposes image-derived data upon the operative scene. We create a model of relevant structures (e.g., tumor volume, blood vessels, and nerves) from multimodality preoperative images. By calibrating microscope optics, registering the patient to image coordinates, and tracking the microscope and patient intraoperatively, we can generate stereo projections of the diree-dimensional model and project diem into the microscope eyepieces, allowing critical structures to be overlaid on the operative scene in die correct position. Measurements with a head phantom gave a root mean square (RMS) error of 1.08 mm, and the estimated error for a human volunteer is between 2 and 3 mm. Initial evaluation in die operating room was very promising. © 1995 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
Publication data is maintained in RPS. Visit https://rps.ucl.ac.uk
 More search options
UCL Researchers
Author
Dept of Computer Science
Author
Dept of Med Phys & Biomedical Eng
Author
Dept of Med Phys & Biomedical Eng
University College London - Gower Street - London - WC1E 6BT Tel:+44 (0)20 7679 2000

© UCL 1999–2011

Search by