Please report any queries concerning the funding data grouped in the sections named "Externally Awarded" or "Internally Disbursed" (shown on the profile page) to your Research Finance Administrator. Your can find your Research Finance Administrator at https://www.ucl.ac.uk/finance/research/rs-contacts.php by entering your department
Please report any queries concerning the student data shown on the profile page to:

Email: portico-services@ucl.ac.uk

Help Desk: http://www.ucl.ac.uk/ras/portico/helpdesk
Publication Detail
Pacing thresholds of temporary epicardial electrodes: variation with electrode type, time, and epicardial position.
  • Publication Type:
    Journal article
  • Publication Sub Type:
    Comparative Study
  • Authors:
    Kallis P, Batrick N, Bindi F, Mascaro G, Chatzis A, Keogh BE, Parker DJ, Treasure T
  • Publication date:
    03/1994
  • Pagination:
    623, 626
  • Journal:
    Ann Thorac Surg
  • Volume:
    57
  • Issue:
    3
  • Status:
    Published
  • Country:
    Netherlands
  • Print ISSN:
    0003-4975
  • PII:
    0003-4975(94)90555-X
  • Language:
    eng
  • Keywords:
    Cardiac Pacing, Artificial, Electric Stimulation, Electrodes, Implanted, Heart Arrest, Induced, Heart Atria, Heart Ventricles, Humans, Pacemaker, Artificial, Pericardium, Time Factors
Abstract
We compared the variation in pacing thresholds of two widely used temporary pacing electrodes at different epicardial sites in 67 patients after coronary artery operations performed with either cardioplegia or ventricular fibrillation. In 33 patients, a bare, braided pacing wire (DW) was placed on the right ventricle and a Medtronic localized epicardial electrode (MED), on each ventricle. In the other 34 patients, the DW wire was placed on the right atrium and a MED electrode, on each atrium. Pacing thresholds were measured at the time of placement; at 1 hour, 6 hours, and 12 hours postoperatively; and daily for 4 days. The pacing thresholds (mean +/- standard error of the mean) at implantation were as follows: DW wire = 0.93 +/- 0.08 V and MED electrode = 0.63 +/- 0.1 V in the ventricles and DW = 1.28 +/- 0.18 V and MED = 0.65 +/- 0.09 V in the atria. On the fourth postoperative day, the pacing thresholds were DW = 2.08 +/- 0.21 V and MED = 1.19 +/- 0.22 V in the ventricles and DW = 2.33 +/- 0.29 V and MED = 1.04 +/- 0.09 V in the atria. The pacing thresholds of both types of wire increased significantly over time, but this deterioration was more pronounced with the braided wire both on the ventricle and on the atrium. The pacing threshold patterns were not affected by chamber side or mode of myocardial preservation. The braided ventricular wire failed to capture after 24 hours in 9 of 30 patients, whereas the localized epicardial electrode captured in all instances (p < 0.001).(ABSTRACT TRUNCATED AT 250 WORDS)
Publication data is maintained in RPS. Visit https://rps.ucl.ac.uk