Please report any queries concerning the funding data grouped in the sections named "Externally Awarded" or "Internally Disbursed" (shown on the profile page) to your Research Finance Administrator. Your can find your Research Finance Administrator at http://www.ucl.ac.uk/finance/research/post_award/post_award_contacts.php by entering your department
Please report any queries concerning the student data shown on the profile page to:

Email: portico-services@ucl.ac.uk

Help Desk: http://www.ucl.ac.uk/ras/portico/helpdesk
Publication Detail
Safety of multi-channel stimulation implants: a single blocking capacitor per channel is not sufficient after single-fault failure.
  • Publication Type:
    Journal article
  • Publication Sub Type:
    Journal Article
  • Authors:
    Nonclercq A, Lonys L, Vanhoestenberghe A, Demosthenous A, Donaldson N
  • Publication date:
    04/2012
  • Pagination:
    403, 410
  • Journal:
    Med Biol Eng Comput
  • Volume:
    50
  • Issue:
    4
  • Status:
    Published
  • Country:
    United States
  • Language:
    eng
  • Keywords:
    Electric Stimulation, Electrodes, Implanted, Electronics, Medical, Equipment Design, Equipment Failure, Equipment Safety, Humans, Prostheses and Implants
Abstract
One reason given for placing capacitors in series with stimulation electrodes is that they prevent direct current flow and therefore tissue damage under fault conditions. We show that this is not true for multiplexed multi-channel stimulators with one capacitor per channel. A test bench of two stimulation channels, two stimulation tripoles and a saline bath was used to measure the direct current flowing through the electrodes under two different single fault conditions. The electrodes were passively discharged between stimulation pulses. For the particular condition used (16 mA, 1 ms stimulation pulse at 20 Hz with electrodes placed 5 cm apart), the current ranged from 38 to 326 μA depending on the type of fault. The variation of the fault current with time, stimulation amplitude, stimulation frequency and distance between the electrodes is given. Possible additional methods to improve safety are discussed.
Publication data is maintained in RPS. Visit https://rps.ucl.ac.uk
 More search options
UCL Researchers
Author
Dept of Electronic & Electrical Eng
Author
Dept of Med Phys & Biomedical Eng
Author
Department of Materials & Tissue
University College London - Gower Street - London - WC1E 6BT Tel:+44 (0)20 7679 2000

© UCL 1999–2011

Search by