Please report any queries concerning the funding data grouped in the sections named "Externally Awarded" or "Internally Disbursed" (shown on the profile page) to your Research Finance Administrator. Your can find your Research Finance Administrator at https://www.ucl.ac.uk/finance/research/rs-contacts.php by entering your department
Please report any queries concerning the student data shown on the profile page to:

Email: portico-services@ucl.ac.uk

Help Desk: http://www.ucl.ac.uk/ras/portico/helpdesk
Publication Detail
Nonrigid registration using free-form deformations: application to breast MR images
  • Publication Type:
    Journal article
  • Publication Sub Type:
    Article
  • Authors:
    Rueckert D, Sonoda LI, Hayes C, Hill DL, Leach MO, Hawkes DJ
  • Publication date:
    08/1999
  • Pagination:
    712, 721
  • Journal:
    IEEE TRANS MED IMAGING
  • Volume:
    18
  • Issue:
    278-0062 (Print), 8
  • Keywords:
    Breast, anatomy & histology, Female, Humans, Image Processing, Computer-Assisted, methods, Magnetic Resonance Imaging, Research Support, Non-U.S.Gov't
  • Addresses:
    Division of Radiological Sciences and Medical Engineering, Guy's, King's, and St. Thomas' School of Medicine, King's College London, Guy's Hospital, London, UK. D.Rueckert@umds.ac.uk
  • Notes:
    DA - 19991115
Abstract
In this paper we present a new approach for the nonrigid registration of contrast-enhanced breast MRI. A hierarchical transformation model of the motion of the breast has been developed. The global motion of the breast is modeled by an affine transformation while the local breast motion is described by a free-form deformation (FFD) based on B-splines. Normalized mutual information is used as a voxel-based similarity measure which is insensitive to intensity changes as a result of the contrast enhancement. Registration is achieved by minimizing a cost function, which represents a combination of the cost associated with the smoothness of the transformation and the cost associated with the image similarity. The algorithm has been applied to the fully automated registration of three-dimensional (3-D) breast MRI in volunteers and patients. In particular, we have compared the results of the proposed nonrigid registration algorithm to those obtained using rigid and affine registration techniques. The results clearly indicate that the nonrigid registration algorithm is much better able to recover the motion and deformation of the breast than rigid or affine registration algorithms