Please report any queries concerning the funding data grouped in the sections named "Externally Awarded" or "Internally Disbursed" (shown on the profile page) to your Research Finance Administrator. Your can find your Research Finance Administrator at http://www.ucl.ac.uk/finance/research/post_award/post_award_contacts.php by entering your department
Please report any queries concerning the student data shown on the profile page to:

Email: portico-services@ucl.ac.uk

Help Desk: http://www.ucl.ac.uk/ras/portico/helpdesk
Publication Detail
Towards a closed-loop transmitter system with integrated class-D amplifier for coupling-insensitive powering of implants
Abstract
The efficiency of an inductive power transfer link is greatly affected by misalignment between the coils, caused by the patient's movement, after implantation. In this paper we present the preliminary design of a coupling-insensitive system with an integrated class-D transmitter, suitable for powering epidural stimulation implants. The operation of the system is based on closed-loop modulation of the power delivered to the implant, to compensate for losses of coupling between coils after implantation. An integrated transmitter with efficiencies above 90% was designed to operate between 1 MHz and 14 MHz. The circuit design employs 0.18-um high-voltage CMOS technology. Simulations are presented to verify the operation of the circuit. © 2012 IEEE.
Publication data is maintained in RPS. Visit https://rps.ucl.ac.uk
 More search options
UCL Researchers
Author
Dept of Electronic & Electrical Eng
Author
Dept of Med Phys & Biomedical Eng
Author
Dept of Electronic & Electrical Eng
University College London - Gower Street - London - WC1E 6BT Tel:+44 (0)20 7679 2000

© UCL 1999–2011

Search by